Productos->PCI->14 y 16 de resolución->PCI-DAS6052

PCI-DAS6052

PCI-DAS6052
16-channel, 333 kS/s analog I/O board with two 16-bit analog outputs, eight digital I/O, and two counter/timers
 
  • Functionally equivalent to National Instruments PCI-6052E
  • 16 channels of 16-bit A/D resolution
  • 333 kS/s maximum sample rate
  • Two 333 kHz 16-bit D/A channels
  • 8 digital I/O bits and 2 counter/timers
  • Triggering of measurement and control via both analog and digital signals
  • Easy synchronization of multiple measurement boards

General Description

The PCI-DAS6052 from Measurement Computing is an accurate and powerful measurement instrument that plugs into any PCI slot in your computer. Combined with powerful software, it turns your personal computer into a measurement system that may be used to automate experiments, construct product test stands, monitor and control production equipment or be embedded in products such as airport security systems and jet fighter test simulators.

Measurements of up to 16 channels at a resolution of 16 bits, (one part in 65,536) on any of 15 inputs ranges may be taken at rates up to 333,000 samples per second. Dual 16-bit analog outputs may be updated at rates to 333 kHz per channel. Discrete digital bi-directional I/O lines (8) can control relays and solenoids, and monitor switch and contact closures.

Analog Input Accuracy
333 kS/s sampling rate, single channel operation and a 15-minute warm-up. Accuracies listed are for measurements made following an internal calibration. They are valid for operational temperatures within ±1°C of internal calibration temperature and ±10°C of factory calibration temperature. Calibrator test source high side tied to Channel 0 High and low side tied to Channel 0 Low. Low-level ground is tied to Channel 0 Low at the user connector.

 
Table 1 - Absolute Accuracy
Range Absolute Accuracy
±10V ±15.6 LSB
±5V ±5.7 LSB
±2.5V ±15.6 LSB
±1V ±15.7 LSB
±500mV ±15.9 LSB
±250mV ±18.0 LSB
±100mV ±21.0 LSB
±50mV ±23.0 LSB
0 to 10V ±8.1 LSB
0 to 5V ±27.8 LSB
0 to 2V ±28.0 LSB
0 to 1V ±28.0 LSB
0 to 500mV ±31.7 LSB
0 to 200mV ±36.4 LSB
0 to 100mV ±38.7 LSB
 
Table 2 - Absolute Accuracy Components - All values are (±)
Range % of Reading Offset (µV) Averaged Noise + Quantization (µV) Temp Drift(%/DegC) Absolute Accuracy at FS (mV)
Single Pt Averaged¹
±10V 0.0371 947 981 87.0 0.0006 4.747
±5V 0.0071 476 491 43.5 0.0001 0.876
±2.5V 0.0371 241 245 21.7 0.0006 1.190
±1V 0.0371 99.2 98.1 8.7 0.0006 0.479
±500mV 0.0371 52.1 56.2 5.0 0.0006 0.243
±250mV 0.0421 28.6 32.8 3.0 0.0006 0.137
±100mV 0.0471 14.4 22.4 2.1 0.0006 0.064
±50mV 0.0471 9.7 19.9 1.9 0.0006 0.035
0 to 10V 0.0071 476 491 43.5 0.0001 1.232
0 to 5V 0.0371 241 245 21.7 0.0006 2.119
0 to 2V 0.0371 99.2 98.1 8.7 0.0006 0.850
0 to 1V 0.0371 52.1 56.2 5.0 0.0006 0.428
0 to 500mV 0.0421 28.6 39.8 3.0 0.0006 0.242
0 to 200mV 0.0471 14.4 22.4 2.1 0.0006 0.111
0 to 100mV 0.0471 9.7 19.9 1.9 .0006 0.059
 
1. Averaged measurements assume dithering and averaging of 100 single-channel readings
 
Table 3 - Differential non-linearity
Range Relative Accuracy (µV)
Single Point Averaged ¹
±10V 1145 115
±5V 573 57.3
±2.5V 286 28.6
±1V 115 11.5
±500mV 66.3 6.6
±250mV 39.2 3.9
±100mV 27.7 2.8
±50mV 25.3 2.5
0 to 10V 66.3 6.6
0 to 5V 286 28.6
0 to 2V 115 11.5
0 to 1V 66.3 6.6
0 to 500mV 48.2 3.9
0 to 200mV 27.7 2.8
0 to 100mV 25.3 2.5
 
General Accessories
The I/O board used in the accessory photos is not a PCI-DAS6052. It is a representative board with the same connector.
 
Cable/Screw Terminal Options

Cable/BNC Terminal Options

External Signal Conditioning